| ====== Freezer monitoring system ====== | ====== Freezer monitoring system ====== | ||
| - | {{ : | + | One of the first projects I did a couple years ago, was designing a custom freezer monitoring system. It's end purpose was monitoring -80°C freezer, but I tested them successfully in liquid nitrogen, down to -196°C. |
| - | One of the first projects I did a couple years ago, was designing a custom freezer monitoring system. It's end purpose was monitoring -80°C freezer, but they where actually tested down to -196°C in liquid nitrogen. | + | {{ : |
| - | The system communicates wirelessly with two gateways, by using a Zigbee mesh network between multiple nodes. | + | The system communicates wirelessly with two gateways, by using a Zigbee mesh network between multiple nodes. |
| - | \\ | + | Some of the collected live data is available at: [[https:// |
| - | \\ | + | |
| - | \\ | + | |
| - | \\ | + | |
| - | \\ | + | |
| - | \\ | + | |
| - | \\ | + | |
| \\ | \\ | ||
| \\ | \\ | ||
| Line 31: | Line 25: | ||
| Optionally it's also possible to connect the module to a freezers alert port, which is basically a relay that triggers whenever the Freezer itself detects an issue. This provides an early warning, even before the temperature inside the freezer rises to much. | Optionally it's also possible to connect the module to a freezers alert port, which is basically a relay that triggers whenever the Freezer itself detects an issue. This provides an early warning, even before the temperature inside the freezer rises to much. | ||
| + | ===== 3D printing: ===== | ||
| + | |||
| + | {{ : | ||
| + | |||
| + | As this system is utilized in a very busy environment, | ||
| + | |||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | \\ | ||
| + | |||
| + | ===== Software and GUI ===== | ||
| + | |||
| + | {{ : | ||
| + | |||
| + | I decided to use a mySQL database for data storage, combined with Grafana as GUI. I'm however in the process of merging all data in an Influx database for performance improvements. | ||
xnorit/public/summary/freezermonitoringsystem.1728820179.txt.gz · Last modified: by timdv91